Performance Verification
Absolute accur
acy at calibration point,
Preamp ON Low limit Test Result High limit
Absolute amplitude SP 41 MHz, RF Path
–0.5 dB +0.5 dB
Absolute amplitude SP 300 kHz, LF Path
–0.5 dB +0.5 dB
Absolute amplitude SP 1 MHz, LF Path
–0.5 dB +0.5 dB
Third Order Intermodulation Distortion Test Result High limit
2.130 GHz
–82 dBc
Displayed Average Noise Level (DANL) Test Result High limit
Normalize
d to 1 Hz RBW, with log-average detector
100 Hz –2 kH
z(LF
Band)
–124 dBm/H
z
2kHz–10kHz(LF
Band)
–141 dBm/Hz
10 kHz 32 MHz (LF
Band)
–150 dBm/Hz
9 kHz 1 MHz
(RSA51
03B/5106B)
–108 dB
m/Hz
1MHz–10MHz
–136 dB
m/Hz
10 MHz
–2GHz
(RSA5103B/5106B)
–154 d
Bm/Hz
10MHz–2GHz
(RSA5115B/5126B)
–152 dBm/Hz
2 GHz 3 GHz –152 dBm/Hz
3GHz–4GHz
(RS
A5106B/5115B/5126B)
–152 dBm/Hz
4G
Hz 6.2 GHz
(RSA5106B/5115B/5126B)
–1
49 dBm/Hz
6.2GHz–13GHz
(RSA5115B/5126B)
–146 dBm/Hz
13 GHz 23 GHz
(
RSA5115B/5126B)
–144 dBm/Hz
23 GHz 26.5 GHz
(RSA5126B)
–140 dBm/Hz
136 RSA5100B Series Technical Reference

Table of Contents

Other manuals for Tektronix RSA5100B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix RSA5100B Series and is the answer not in the manual?

Tektronix RSA5100B Series Specifications

General IconGeneral
BrandTektronix
ModelRSA5100B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals