EasyManua.ls Logo

Tektronix T922 - Page 76

Tektronix T922
217 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Service Information-T921/T922/T922R
REMOVE
2
SCREWS
TO
REMOVE
HANDLE
SLIDE
VERTICAL
MODULE
TO
REAR
AND
LIFT
TO
REMOVE
ONE
SCREW
AND
REMOVE
THREE
3/16
INCH
POSTS
.
Fig
.
5-10
.
T922R
vertical
module
replacement
.
2
.
Make
note of
their
location
to
aid
in
reassembly,
then
disconnect
:
P2
from
horizontal
module
P2412
from
buffer
board
P2103
from
EXT
TRIG
switch
board
(3
wire
connec-
tor)
Thetwo
horizontal
deflection
leads
3
.
Remove
one
screw
and
three
3/16-inch posts
(see
Fig
.
5-11)
.
Fig
.
5-11
.
T922R
horizontal
module
replacement
.
2
.
Remove
the
knob/shafts
and
shafts
from
the
follow-
ing
controls
:
INTENSITY
BEAM
FIND
FOCUS
POWER
WARNING
To
avoid
a
shock
hazard,
be
certain
you
have
disconnected
the
power
cord
before continuing
.
3
.
Remove
the
four
screws
from
the
high-voltage
shield
and
lift
the
shield
away
from
the
instrument
.
4
.
Make
note
of
the
line
cord
color
code
and
position, to
aid
in
reassembly
.
Then,
unsolder
the
line
cord
conduc-
tors
from
the
interface
board
.
5
.
Push
the
clear
plastic
POWER
light-pipe
out
the
front
of
the
instrument
to
remove
.
WARNING
4
.
Lift
horizontal
module
away
from
instrument
to
The
crt
anode
and
the
high
voltage
multiplier
will
remove
.
hold a
charge
after
you
turn
off the
instrument
.
To
prevent
a
shock
hazard,
use
only
one
hand
to
remove
the
anode
lead
.
Then,
touch
the
anode
lead
to
one
of
the
posts
to
discharge
the
crt
(see
Fig
.
5-
12)
.
Also
take
a short
piece
of
insulated
wire
and
Interface
Board
Replacement
short the
outpput
of the
high-voltage
multiplier
to
the chassis (see
Fig
.
5-12)
.
1
.
Remove
four
screws
from
the
left
side
rail
and
remove
the
side
rail
.
6
.
Pull
the
crt
anode
lead
forward
to
remove
.
5-
1
8

Related product manuals