Service
Information-T921/T922/T922R
WARNING
The
crt
anode
and
the
high-voltage
multiplier
will
hold
a
charge
after
you
turn
off
the
instrument
.
To
prevent
a
shock
hazard,
use
only
one
hand
to
remove
the
anode
lead
.
Then,
touch
the
anode
lead
to
one
of the
posts
to
discharge
the
crt
(see
Fig
.
5-
12)
.
Also,
take a short
piece
of
insulated
wire
and
short
the
output
of
the
high-voltage
multiplier
to
the
chassis
(see
Fig
.
5-12)
.
3
.
Slide
the
crt
anode
lead
out
from
under
the
plastic
clip
which
holds
it
next
to
the
chassis near the
interface
board
.
4
.
Make
note
of
the
positions
of
the
four
crt
leads
to
aid
in
reassembly
.
Remove
the
four
crt
leads
from
the
neck
of
the
crt
.
5
.
Remove
crt
socket
from
the
base
of
the
crt
.
6
.
Lift
the
plastic
mask
and
light
filter
away
from
the
crt
to
remove
.
After
you
remove
the
bezel,
the
crt
can
slide
forward
.
To
avoid
crt
damage,
hold
one
hand
over
the
crt
face
if
you
must
turn
or
move
the instrument
.
7
.
Remove
the
four
screws
holding
the
bezel
.
Lift
the
bezel
and
implosion
shield
away
from
the
crt
.
8
.
Remove
the
graticule
illumination
assembly
.
9
.
While
holding
one
hand
over the
crt
face,
push
the
base
of
the
crt
forward
.
When
you
are
able,
firmly
grasp
the
front
of
the
crt
and
slide
it
out
of
the
crt
shield
assembly
.
SCALE
ILLUM
Lamp
Replacement
1
.
Lift
the
plastic
mask
and
light
filter
away
from
crt
to
remove
.
2
.
Disconnect
P816
from
the
SCALE
ILLUM
assembly
.
3
.
Pull
SCALE
ILLUM
assembly
away
from
the
crt
to
remove
.
If
this
is
difficult,
push
against
the
two
screws
from
behind
the
front
panel
(see
Fig
.
5-14)
.
4
.
Remove
the
two
screws
holding the
plastic
piece
to
the
circuit
board
.
Remove
the
plastic
piece
.
5
.
Unsolder
the
defective
lamp
and
replace
.
Fuse
Replacement
WARNING
Line
voltage
is
present
on
the
fuse
whenever
the
power
cord
is
plugged
in
.
To
prevent
a
shock
hazard,
disconnect
the
power
cord
.
The
fuse
is
located
in
a
holder
near
the
rear
of
the
interface
board
.
To
remove,
grasp
firmly
and
pull
away
from
the
interface
board
.
Fig
.
5-14
.
T922R
scale
illum
assembly
replacement
.