Aoucndix B: Performance Verification
Test Equipment
These procedures use external, traceable signal sources to directly check
instrument perfotmancc. If your test equipment does, not meet the minimum
requirements listed in Table B-l, your test results will be invalid.
Table B-i: Test equipment
Item number
and description Minimum requirements Example Purpose
1. Termination 50 0
Impedance 50 R; connectors: Tektronix pari number Checking
delay belween
(two required) female ENC inpuf, male BNC 011-0049-01
channels
2. Cable, Precision Coaxial
(Iwo required)
3. Connector,
D&Banana
50 0.91 cm (36 in), male to
male ENC connectors
Female-BNC to dual-banana
Tektronix part number
012-0462-00
Tektronix part number
103-0090-00
Signal interconnec!:ion
Several accuracy tests
4. Connector, Male-BNC to dual-female-BNC Tektronix part number Checking trigger sensitivity
BNCT 103-0030-00
5. Coupler, Female-ENC to dual-male-BNC Tektronix part number Checking delay
between
Dual-Input
0670525-02
channels
6. Generator, DC Calibration Vanable amplitude to f110 V;
Wavetek 9100 Calibration Checking DC offset, gain, and
accurecy to 0.1% System with Oplion 250 measurement accuracy
7. Generator, 200 kHz to 250 MHz: variable
Wavetek 9100 Calibration Checking bandwidth and trigger
Leveled Sine Wave,
amplitude from 5 mV to 4 Vpp
System with Option 250
sensitivity
Medium-Frequency
into 50 R
6. Generator, 200 kHz to 400 MHz; variable
Rohde & Schwarr SMY with
Checking bandwidth and trigger
Leveled Sine Wave, amplitude from 5 mV to 4 V,.,
URV 35 Power Meter and sensitivity
High-Frequency’
into 50 n
NRV-Z8 Power Sensor
9. Generator, lme Mark Variable marker frequency from Wavetek 9100 Calibration Checking sample rate and
10 ms to 10 ns; accuracy within System with Option 250 delay-lime accuracy
10. Probe, 10X.
A P6109B (TDS &IO), P6111B Tektronix number
Signal interconnection
included with this FDS 360), or P6114E P6109B(TDS340),P6111E
instrument (TDS 360) probe
(TDS 360). or P61148
FDS 380)
’ The high frequency leveled sine wave generator is only required to verify the TDS 380, not the TOS 340 or TDS 360. If you
use the example equipment, refer to
Sine
Wave Generator
Leveling
Procedure on page E-14 for information on obtaining
a leveled output from an unleveled sine wave generator, If available, you can use a Tektronix 56504 Leveled Sine Wave
Generatior in place of the example equipment.
TDS 340, TDS 360 & ‘TDS 380 User Manual
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com