Performance Tests
1–20
TDS 500B, TDS 600B & TDS 700A Performance Verification and Specifications
TDS 600B Test Record (Cont.)
Instrument Serial Number: Certificate Number:
Temperature: RH %:
Date of Calibration: Technician:
TDS 600B Performance Test MaximumOutgoingIncomingMinimum
CH2 1 V Vert scale setting,
–5 Div position setting, +10 V offset
+ 17.2425 V __________ __________ + 18.7575 V
CH2 1 V Vert scale setting,
+5 Div position setting, –10 V offset
– 18.7575 V __________ __________ – 17.2425 V
CH3 5 mV Vert scale setting,
or AX1 –5 Div position setting, +1 V offset
+ 1.0329 V __________ __________ + 1.0471 V
CH3 5 mV Vert scale setting,
or AX1 +5 Div position setting, –1 V offset
– 1.0471 V __________ __________ – 1.0329V
CH3 200 mV Vert scale setting,
or AX1 –5 Div position setting, +10 V offset
+ 11.4165 V __________ __________ + 11.7835 V
CH3 200 mV Vert scale setting,
or AX1 +5 Div position setting, –10 V offset
– 11.7835 V __________ __________ – 11.4165 V
CH3 1 V Vert scale setting,
or AX1 –5 Div position setting, +10 V offset
+ 17.2425 V __________ __________ + 18.7575 V
CH3 1 V Vert scale setting,
or AX1 +5 Div position setting, –10 V offset
– 18.7575 V __________ __________ – 17.2425 V
CH4 5 mV Vert scale setting,
or AX2 –5 Div position setting, +1 V offset
+ 1.0329 V __________ __________ + 1.0471 V
CH4 5 mV Vert scale setting,
or AX2 +5 Div position setting, –1 V offset
– 1.0471 V __________ __________ – 1.0329V
CH4 200 mV Vert scale setting,
or AX2 –5 Div position setting, +10 V offset
+ 11.4165 V __________ __________ + 11.7835 V
CH4 200 mV Vert scale setting,
or AX2 +5 Div position setting, –10 V offset
– 11.7835 V __________ __________ – 11.4165 V
CH4 1 V Vert scale setting,
or AX2 –5 Div position setting, +10 V offset
+ 17.2425 V __________ __________ + 18.7575 V
CH4 1 V Vert scale setting,
or AX2 +5 Div position setting, –10 V offset
– 18.7575 V __________ __________ – 17.2425 V
Analog Bandwidth
CH1 100 mV 424 mV __________ __________ N/A
CH2 100 mV 424 mV __________ __________ N/A
CH3 100 mV
or AX1
424 mV __________ __________ N/A
CH4 100 mV
or AX2
424 mV __________ __________ N/A