Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–113
H DESE 128
H *ESE 128
H *SRE 32
H *PSC 0
DIAg:STATE EXECUTE
executes all the diagnostic tests that have been selected.
DISplay? (Query Only)
Returns the current display settings.
Display
DISplay?
DISplay
?
DISPLAY?
might return :DISPLAY:FORMAT YT;STYLE VECTORS;FILTER SINX;PERSISTĆ
ENCE 500.0E-3;GRATICULE FULL;TRIGT 1;INTENSITY:OVERALL 85;
WAVEFORM 70;TEXT 60;CONTRAST 150
DISplay:CLOCk
Controls the display of the date and time. This is equivalent to setting Display
Date/Time in the Readout Options side menu. The query form returns an ON (1)
or an OFF (0).
Display
DISplay:CLOCk { OFF | ON | <NR1> }
DISplay:CLOCk?
Examples