Command Descriptions
2–196
TDS Family Oscilloscope Programmer Manual
MASK:FILTer { ENAbled | DISAbled }
MASK:FILTER?
<Space>
?
ENAbled
DISAbled
FILTer
MASK
:
ENAbled enables the digital filter. This is the default value.
DISAbled disables the digital filter.
MASK:FILTER ENABLED
enables the digital filter.
MASK:MASK<n>? (Query Only)
TDS 500C & 700C Only
Returns the state of all settable parameters of the specified mask.
Mask
MASK:MASK<NR1>?
MASK
?
<NR1>
MASK
:
<NR1> is 1 to 8.
MASK:MASK1?
might return
:MASK:MASK1:POINTSP
11.400000000E+0,12.531328321E+0,8.820e+01,1.253e+01,8.840e+01,0.0
E+0,1.120e+01,0.0E+0
MASK:MASK<n>
TDS 500C & 700C Only
Deletes all points in the specified mask.