Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–289
TRIGGER:MAIN:LEVEL 0.5
sets the main trigger level to 0.5 V.
TRIGger:MAIn:LOGIc? (Query Only)
Not on TDS 400A
Returns all main logic trigger parameters.
Trigger
TRIGger:MAIn:LOGIc?
:
MAIn
TRIGger :
?
LOGIc
TRIGGER:MAIN:LOGIC?
might return :TRIGGER:MAIN:LOGIC:CLASS PATTERN;
FUNCTION AND;WHEN TRUE;THRESHOLD:CH1 0;CH2 0;CH3 0;
CH4 0;:TRIGGER:MAIN:LOGIC:INPUT:CH1 HIGH;CH2 X;
CH3 X;:TRIGGER:MAIN:LOGIC:PATTERN:INPUT:CH4 X;
:TRIGGER:MAIN:LOGIC:STATE:INPUT:CH4 RISE
TRIGger:MAIn:LOGIc:CLAss
Not on TDS 400A
Sets or queries the type of main logic trigger. This command is equivalent to
selecting Class in the Trigger menu when the Type is set to Logic.
Trigger
TRIGger:MAIn:LOGIc:CLAss { PATtern | STATE
SETHold (TDS 500C, 600B & 700C only)}
TRIGger:MAIn:LOGIc:CLAss?
Group
Syntax
Examples
Group
Syntax

Table of Contents