Command Descriptions
2–56
TDS Family Oscilloscope Programmer Manual
NOTE. If you set the acquisition system to single sequence, envelope mode, and
set the number of envelopes to infinity, the digitizing oscilloscope will envelope a
maximum of 2001 acquisitions.
ACQUIRE:STOPAFTER RUNSTop
sets the oscilloscope to stop acquisition when the user presses the front-panel
RUN/STOP button.
ACQUIRE:STOPAFTER?
might return SEQUENCE.
ALIas
Turns command aliases on or off. This command is identical to the
ALIas:STATE command.
Alias
ALIas { OFF | ON | <NR1> }
ALIas?
ON
ALIas
<Space>
?
<NR1>
OFF
OFF or <NR1> = 0 turns alias expansion off. If a defined alias label is sent when
ALIas is OFF, an execution error (110, “Command header error”) will be
generated.
ON or <NR1> 0 turns alias expansion on. When a defined alias is received, the
specified command sequence is substituted for the alias and executed.
ALIAS ON
turns the alias feature on.
ALIAS?
returns 1 when aliases are on.
Examples
Group
Syntax
Arguments
Examples

Table of Contents

Other manuals for Tektronix TDS 500C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS 500C and is the answer not in the manual?

Tektronix TDS 500C Specifications

General IconGeneral
BrandTektronix
ModelTDS 500C
CategoryTest Equipment
LanguageEnglish

Related product manuals