Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–75
POSition
CH
<x>
:
?
<Space> <NR3>
<NR3> is the desired position, in divisions from the center graticule. The range is
±5 divisions.
CH2:POSITION 1.3E+00
positions the channel 2 input signal 1.3 divisions above the center of the display.
CH1:POSITION?
might return -1.3E+00, indicating that the current position of channel 1 is at
–1.3 divisions.
CH<x>:PROBE? (Query Only)
Returns the attenuation factor of the probe that is attached to the specified
channel.
Vertical
CH<x>:PROBE?
CH
<x> :
PROBE
?
<NR3>
CH4:PROBE?
might return 100.0E-3 for a 10x probe.
CH<x>:PROBECal? (Query Only)
TDS 500C, 600B, & 700C
Queries the probe cal status for the selected channel.
Vertical
CH<x>:PROBECal?
Arguments
Examples
Group
Syntax
Returns
Examples
Group
Syntax