Nominal Traits
TDS 500D, TDS 600B & TDS 700D Performance Verification and Specifications
2–13
Table 2–4: Nominal traits — Time base system (cont.)
Name Description
Record Length Selection TDS 500D & TDS 700D:
500, 1,000, 2,500, 5,000, 15,000 and 50,000 samples
In addition to the record lengths previously listed, the following record lengths are
available with the following options:
TDS 520D & TDS 724D with option 1M:
1-channel: up to 250,000 samples
2-channels: up to 130,000 samples
TDS 520D & TDS 724D with option 2M:
1-channel: up to 4,000,000 samples
2-channels: up to 2,000,000 samples
TDS 540D, TDS 580D, TDS 754D, TDS 784D & TDS 794D with option 1M:
1-channel: up to 500,000 samples
2-channels: up to 250,000 samples
3 or 4-channels up to 130,000 samples
TDS 540D, TDS 580D, TDS 754D, TDS 784D & TDS 794D with option 2M:
1-channel: up to 8,000,000 samples
2-channels: up to 4,000,000 samples
3 or 4-channels up to 2,000,000 samples
1
The range of real-time rates, expressed in samples/second, at which a digitizer samples signals at its inputs and stores
the samples in memory to produce a record of time-sequential samples.
2
The range of waveform rates for interpolated (or equivalent-time on the TDS 700D) waveform records.
3
The Waveform Rate (WR) is the equivalent sample rate of a waveform record. For a waveform record acquired by
real-time sampling of a single acquisition, the waveform rate is the same as the real-time sample rate; for a waveform
created by interpolation of real-time samples from a single acquisition or, on applicable products, the equivalent-time
sampling of multiple acquisitions, the waveform rate created is faster than the real time sample rate. For all these cases,
the waveform rate is 1/(Waveform Interval) for the waveform record, where the waveform interval (WI) is the time between
the samples in the waveform record.
Table 2–5: Nominal traits — Triggering system
Name Description
Range, Delayed Trigger Time Delay 16 ns to 250 s
Range, Events Delay TDS 600B; 2 to 10,000,000
TDS 500D/700D: 1 to 10,000,000
Range (Time) for Pulse-Glitch,
Pulse-Width, Time-Qualified Runt, Timeout,
or Slew Rate Trigger, Delta Time
1 ns to 1 s