EasyManua.ls Logo

Tektronix TDS 500D series - Figure 1-17: Measurement of Trigger-Level Accuracy

Tektronix TDS 500D series
154 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Tests
TDS 500D, TDS 600B & TDS 700D Performance Verification and Specifications
1–55
Set DC calibration generator
to equal offset (+10 V).
2
Set vertical position to –3 divs.
Set vertical offset to +10 volts.
1
Read results.
3
Figure 1–17: Measurement of trigger-level accuracy
3. Confirm Delayed trigger system is within limits for Trigger-level/Threshold
accuracy:
a. Select the Delayed time base:
H Press HORIZONTAL MENU.
H Press the main-menu button Time Base.
H Press the side-menu buttons Delayed Only and Delayed Trigger-
able.
H Set D (delayed) horizontal SCALE to 500 s.
b. Select the Delayed trigger system:
H Press SHIFT; then press the front-panel DELAYED TRIG button.
H Press the main-menu button Level.
c. Measure the test signal: Press the side-menu button SET TO 50%.
Read the measurement results in the side (or main) menu below the label
Level.

Table of Contents

Other manuals for Tektronix TDS 500D series

Related product manuals