Performance Tests
1–48
TDS 500D, TDS 600B & TDS 700D Performance Verification and Specifications
Check long–term sample
rates and delay time
accuracies against limits.
4
Align the trigger T to the
center graticule line.
1
Set
horizontal
mode.
2
Set horizontal scale
and delayed time.
3
Figure 1–13: Measurement of accuracy – Long-term and delay time
3. Disconnect the hookup: Disconnect the cable from the generator output at
the input connector of CH 1.