EasyManua.ls Logo

Tektronix TDS 500D series - Figure 1-25: Measurement of Probe Compensator Amplitude

Tektronix TDS 500D series
154 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Tests
1–70
TDS 500D, TDS 600B & TDS 700D Performance Verification and Specifications
Figure 1–25: Measurement of probe compensator amplitude
g. Check against limits:
H Subtract the value just obtained (base level) from that obtained
previously (top level).
H CHECK that the difference obtained is within 495 mV to 505 mV,
inclusive.
H Enter voltage difference on test record.
3. Disconnect the hookup: Disconnect the cable from CH 1.

Table of Contents

Other manuals for Tektronix TDS 500D series

Related product manuals