Performance Tests
1–82
TDS 500D, TDS 600B & TDS 700D Performance Verification and Specifications
Figure 1–37: Line count accuracy correct result waveform
H Disconnect all test equipment (TSG 121) from the oscilloscope.
H Turn off cursors by pressing CURSOR, then the main-menu
Function button, and, finally, Off from the side menu.
7. Check the Sync Duty Cycle.
a. Set up oscilloscope for Sync Duty Cycle Test:
H Press TRIGGER MENU.
H Press the Standard pop-up to select FlexFmt. Trigger Type should
already be set to Video.
H Press the main-menu Setup.
H Press the side-menu Field Rate.
H Use the keypad to set the field rate to 60.05 Hz (press 60.05, then
ENTER).
H Press the side-menu Lines.
H Use the keypad to set the field rate to 793 lines (press 793, then
ENTER).
H Press the side-menu Fields.