EasyManuals Logo

Tektronix TDS 520A User Manual

Tektronix TDS 520A
291 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #259 background imageLoading...
Page #259 background image
Appendix B: Algorithms
Appendices
AĆ22
For example, if

is set directly, then

would not change even if
samples were out of range. However, if

was chosen using the %
choice from the Set Levels in % Units selection of the Measure menu, then

could give a “CLIPPING” warning.
NOTE
When measurements are displayed using Snapshot, out of range
warnings are NOT available. However, if you question the validity of
any measurement in the snapshot display, you can select and
display the measurement individually and then check for a warning
message.

Other manuals for Tektronix TDS 520A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS 520A and is the answer not in the manual?

Tektronix TDS 520A Specifications

General IconGeneral
BrandTektronix
ModelTDS 520A
CategoryTest Equipment
LanguageEnglish

Related product manuals