Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–87
DATa
Sets or queries the format and location of the waveform data that is transferred
with the CURVe command. Since DATa:DESTination and DATa:TARget are
equivalent, only DATa:DESTination is returned by the DATa? query.
Waveform
CURVE, WAVFrm
DATa { INIT | SNAp }
DATa?
DATa
<Space>
?
INIT
SNAp
INIT initializes the waveform data parameters to their factory defaults.
SNAp sets DATa:STARt and DATa:STOP to match the current vertical bar cursor
positions.
DATA SNAP
assigns the current position of the vertical bar cursors to DATA:START and
DATA:STOP.
DATA?
might return the string :DATA:ENCDG RPBINARY;DESTINATION REF4; SOURCE
REF4;START 1;STOP 500;WIDTH 2
DATa:DESTination
Sets or queries the reference memory location for storing waveform data that is
transferred into the digitizing oscilloscope by the CURVe command. This
command is identical to the DATa:TARget command.
Waveform
Group
Related Commands
Syntax
Arguments
Examples
Group