Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–69
Table 2–27: Commands that Affect BUSY? Response
Operation Command
Single sequence acquisition ACQuire:STATE ON or
ACQuire:STATE RUN
(when ACQuire:STOPAfter is set to SEQuence)
Hardcopy output HARDCopy STARt

might return , indicating that the instrument is busy.
*CAL? (Query Only)
Instructs the digitizing oscilloscope to perform an internal self-calibration and
return its calibration status.
NOTE. The self-calibration can take several minutes to respond. No other
commands will be executed until calibration is complete.
Calibration and Diagnostic


 = 0 indicates that the calibration completed without any errors detected.
 0 indicates that the calibration did not complete successfully.

performs an internal self-calibration and might return to indicate that the
calibration was successful.
CH<x>? (Query Only)
Returns the vertical parameters. Because CH<x>:SCAle and CH<x>:VOLts are
identical, only CH<x>:SCAle is returned.
Examples
Group
Syntax
Returns
Examples

Table of Contents

Other manuals for Tektronix TDS 600B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS 600B and is the answer not in the manual?

Tektronix TDS 600B Specifications

General IconGeneral
BrandTektronix
ModelTDS 600B
CategoryTest Equipment
LanguageEnglish

Related product manuals