Warranted Characteristics
Specifications
1Ć14
TableĂ1Ć8:ăWarranted Characteristics Ċ Signal Acquisition System (Cont.)
Name Description
Analog Bandwidth, DCĆ50Ă
Coupled, or DCĆ1 M Coupled with
P6139A Probe
Volts/Div
5 mV/div - 10 V/div
2 mV/div - 4.98 mV/div
1 mV/div - 1.99 mV/div
Bandwidth
3
DC - 500 MHz
DC - 300 MHz
DC - 200 MHz
Cross Talk (Channel Isolation) 100:1 at 100 MHz and 30:1 at the rated bandwidth for the chanĆ
nel's sensitivity rating, for any two channels having equal volts/diviĆ
sion settings
Input Impedance, DC-1ĂM Coupled 1M±0.5% in parallel with 10 pF ±2 pF
Input Impedance, DC-50Ă Coupled
50 ±1% with VSWR 1.3:1 from DC - 500 MHz
Input Voltage, Maximum, DC-1ĂM,
AC-1ĂM, or GND Coupled
±400 V (DC + peak AC); derate at 20 dB/decade above 1 MHz
Input Voltage, Maximum, DCĆ50Ă or
AC-50Ă Coupled
5V
RMS
, with peaks less than or equal to ±30 V
Lower Frequency Limit, AC Coupled 10 Hz when AC-1ĂM Coupled; 200 kHz when AC-50Ă
Coupled
4
1 To ensure the most accurate measurements possible, run an SPC calibration first. When using the Digitizing Oscilloscope at a
Volts/Div setting of <5 mV/div, an SPC calibration should be run once per week to ensure that instrument performance levels meet
specifications.
2 Net Offset = Offset - (Position x Volts/Div). Net Offset is the voltage level at the center of the AĆD converter's dynamic range. Offset
Accuracy is the accuracy of this Voltage level.
3 The limits given are for the ambient temperature range of 0_Cto+30_C. Reduce the upper bandwidth frequencies by 2.5 MHz for
each _C above +30_C.
4 The AC Coupled Lower Frequency Limits are reduced by a factor of 10, when 10X passive probes are used.
TableĂ1Ć9:ăWarranted Characteristics Ċ Time Base System
Name
Description
Accuracy, Long Term Sample Rate
and Delay Time
±100 ppm over any 1 ms interval