Command Descriptions
2–144
TDS Family Oscilloscope Programmer Manual
sequence (1–2.5–5 on the TDS 620A, 640A, & 644A) will be set to the closest
valid value. If the delayed time base scale is set slower than the main time base
scale, both the main and delayed time base scales will be set to the delay scale
value.
HORIZONTAL:DELAY:SCALE 2.0E-6
sets the delay scale to 2 ms per division.
HORIZONTAL:DELAY:SCALE 9.0E-6
sets the delay scale to 10 ms per division. Since 9 ms is not a valid value within
the 1–2–5 sequence (1–2.5–5 on the TDS 620A, 640A, & 644A), it is automati-
cally set to the closest valid value.
HORIZONTAL:DELAY:SCALE?
might return 1.0E-3, indicating that the delay time is 1 ms per division.
HORizontal:DELay:SECdiv
This command is identical to the HORizontal:DELay:SCAle command. It is
provided to maintain program compatibility with some older models of
Tektronix digitizing oscilloscopes.
Horizontal
HORizontal:DELay:SECdiv <NR3>
HORizontal:DELay:SECdiv?
:
SECdiv
:
DELay
HORizontal
?
<Space> <NR3>
HORizontal:DELay:TIMe
Sets or queries the delay time to wait after the main trigger before the delayed
time base begins. This is equivalent to setting Delayed Runs After Main in the
Time Base side menu of the Horizontal menu.
Horizontal
HORizontal:DELay:MODe
Examples
Group
Syntax
Group
Related Commands