Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–231
TRIGger:MAIn:HOLdoff?
:
MAIn
TRIGger :
?HOLdoff
TRIGGER:MAIN:HOLDOFF?
for the TDS 684A and 7XXA, might return
:TRIGGER:MAIN:HOLDOFF:TIME 250.0E-9;BY DEFAULT
or, for other TDS 4XXA, 5XXA, and 6XXA, might return
:TRIGGER:MAIN:HOLDOFF:VALUE 0.
TRIGger:MAIn:HOLdoff:ACTUal? (Query Only)
TDS 684A and 7XXA Only
Returns the main trigger holdoff value in seconds. This is equivalent to selecting
Mode & Holdoff from the main Trigger menu and viewing the value in the side
menu Holdoff Default or Holdoff Time items (whichever is highlighted).
Trigger
TRIGger:MAIn:HOLdoff:ACTUal?
:
MAIn
TRIGger :
?HOLdoff
ACTUal
:
TRIGGER:MAIN:HOLDOFF:ACTUAL?
might return 4E–6 showing the holdoff time is set to 4 ms.
TRIGger:MAIn:HOLdoff:BY
TDS 684A & 7XXA Only
Sets or queries the main trigger holdoff default. This is equivalent to selecting
Mode & Holdoff from the main Trigger menu, then setting Default Holdoff or
Holdoff (Time) in the resulting side menu.
Trigger
TRIGger:MAIn:HOLdoff:BY { TIMe | DEFAult }