Using Features for Advanced Applications
TDS 500B, TDS 600B, & TDS 700A User Manual
3–145
Figure 3–83: Acquire Menu — Create Limit Test Template
4. Press ±V Limit (side). Enter the vertical (voltage) tolerance value using the
general purpose knob or keypad.
5. Press ±H Limit (side). Enter the horizontal (time) tolerance value using the
general purpose knob or keypad.
Tolerance values are expressed in fractions of a major division. They
represent the amount by which incoming waveform data can deviate without
having exceeded the limits set in the limit test. The range is from 0 (the
incoming waveform must be exactly like the template source) to 5 major
divisions of tolerance.
6. When you have finished specifying the limit test template, press OK Store
Template (side). This action stores the specified waveform in the specified
destination, using the specified tolerances. Until you have done so, the
template waveform has been defined but not created.
To avoid overwriting the template you have just created, store any new
template you create in a different destination from that just stored.
To view the template you have created, press the MORE button. Then press
the button corresponding to the destination reference memory you have used.
The waveform appears on the display.

Table of Contents

Other manuals for Tektronix TDS 700A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS 700A and is the answer not in the manual?

Tektronix TDS 700A Specifications

General IconGeneral
BrandTektronix
ModelTDS 700A
CategoryTest Equipment
LanguageEnglish

Related product manuals