Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–65
CH<x>:POSition
CH<x>:OFFSet <NR3>
CH<x>:OFFSet?
CH
<x> :
OFFSet
?
<Space> <NR3>
<NR3> is the desired offset in volts. The range is dependent on the scale and the
probe attenuation factor. The offset ranges are shown below.
Table 2–25: Offset Ranges for the TDS 400A, 510A, 540B, 600B, 744A, 754A, & 784A
(All Channels) and the TDS 520B, 724A, & 782A (Channel 1 & Channel 2) using a 1x
Probe
CH<x>:SCAle OFFSet Range
1 mV/div – 99.5 mV/div ±1 V
100 mV/div – 995 mV/div ±10 V
1 V/div – 10 V/div ±100 V
Table 2–26: Offset Ranges for the TDS 520B, 724A, & 782A (Aux 1 & Aux 2) using a
1x Probe
CH<x>:SCAle OFFSet Range
50 mV/div & 100 mV/div ±.5 V
500 mV/div & 1 V/div ±5.0 V
5 V/div & 10 V/div ±50 V
CH1:OFFSET 0.5E+00
lowers the channel 1 displayed waveform by 0.5 volts.
CH1:OFFSET?
might return 500.0E-3, indicating that the current channel 1 offset is 0.5 volts.
Related Commands
Syntax
Arguments
Examples