Specifications
TDS1000 and TDS2000 S eries Digital Storage Oscilloscopes Service Manual
1-3
Table 1- 1: Oscilloscope specifications
Vertical
n Analog Bandwidth
TDS1001 TDS1002, TDS2002, TDS2004 TDS1012, TDS2012, TDS2014 TDS2022, TDS2024
n
A
n
a
l
o
g
B
a
n
d
w
i
d
t
h
in Sample and Average
modes at BNC or with
P2200 probe,
DC Coupled
40 MHz{*
60 MHz{* 100 MHz{* 200 MHz{*
0 °Cto+40°C(32°F
to 104 °F)
160 MHz{*
0 °Cto+50°C(32°F
to 122 °F)
20 MHz* (when vertical scal e is set to < 5 mV)
Analog Bandwidth i n TDS1001 TDS1002, TDS2002, TDS2004 TDS1012, TDS2012, TDS2014, TDS2022, TDS2024
g
Peak Detect mode
(
5
0
s
d
i
v
t
o
5
s
d
i
v
**
)
30 MHz{* 50 MHz{* 75 MHz{*
(
5
0
s
d
i
v
t
o
5
s
d
i
v
*
*
)
,
typical
20 MHz* (when vertical scal e is set to < 5 mV)
Selectable Analog Band-
width Limit, typical
20 MHz*
Lower Frequency
Limit, AC Coupled
10 Hz at BNC
1 Hz when using a 10X passive probe
Rise Time at BNC, TDS1001 TDS1002, TDS2002, TDS2004 TDS1012, TDS2012, TDS2014 TDS2022, TDS2024
,
typical
< 8.4ns <5.8ns <3.5ns <2.1ns
Peak Detect Response** Captures 50% or greater amplitude of pulses 12 ns wide typical (50 s/div to 5 s/div) in the center 8
vertical divisions
n DC Gain Accuracy
±3% for Sample or Average acquisition mode, 5 V/div to 10 mV/div
n
D
C
G
a
i
n
A
c
c
u
r
a
c
y
±4% for Sample or Average acquisition mode, 5 mV/div and 2 mV/div
n DC Measurement
Measurement Type Accuracy
n
D
C
M
e
a
s
u
r
e
m
e
n
t
Accuracy, Average
Acquisition Mode
Average of 16 waveforms with vertical
position at zero
±(3% × reading + 0.1 div + 1 mV) when 10 mV/div or greater
is selected
Average of 16 waveforms with vertical
position not at zero
±[3% × (reading + vertical position) + 1% of vertical position
+0.2div]
Add 2 mV for set tings from 2 mV/div t o 200 mV/div
Add 50 mV for set tings from > 200 mV/div to 5 V/div
Volts M easurem ent
Repeatability, Average
Acquisition Mode
Delta volts between any two averages of
16 waveforms acquired under same setup
and ambient condi tions
±(3% × reading + 0.05 div)
{ When vertical scale is set to 5mV.
* Bandwidth reduced to 6 MHz with a 1X probe.
** The oscilloscope reverts to Sample mode when the SEC/DIV (horizontal scale) is set from 2.5 s/div to 5 ns/div on 1 GS/s models,
or from 2.5 s/div to 2.5 ns/div on 2 GS/s models. The Sample mode can still capture 10 ns glitches.

Table of Contents

Other manuals for Tektronix TDS1000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS1000 Series and is the answer not in the manual?

Tektronix TDS1000 Series Specifications

General IconGeneral
BrandTektronix
ModelTDS1000 Series
CategoryTest Equipment
LanguageEnglish