Command Descriptions
2−130
TDS200/1000/2000 Series Oscilloscope Programmer Manual
*RST (No Query Form)
(Reset) Returns the oscilloscope to a known set of oscilloscope
settings, but does not purge any stored settings. This command
executes a subset of the FACtory command.
Status and Error
*RST
Sending the *RST command does the following:
H Returns the oscilloscope settings to the factory defaults (refer to
Appendix B)
H Sets the macro defined by *DDT to a zero-length field
H Clears the pending operation flag and associated operations
The *RST command does not alter the following items:
H The state of the RS-232 or GPIB interface
H Calibration data that affects device specifications
H The selected GPIB oscilloscope address
H Stored settings or waveforms
H The Output queue
H The Service Request Enable Register settings
H The Standard Event Status Enable Register settings
H The Power-On Status Clear flag setting
H Hard copy parameters
H Response Header Enable State
H Front panel LOCK state
H Verbose State
Group
Syntax

Table of Contents

Other manuals for Tektronix TDS1012

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS1012 and is the answer not in the manual?

Tektronix TDS1012 Specifications

General IconGeneral
BrandTektronix
ModelTDS1012
CategoryTest Equipment
LanguageEnglish

Related product manuals