Command Descriptions
TDS200/1000/2000 Series Oscilloscope Programmer Manual
2−131
FACtory, *PSC, *RCL, RECAll:SETUp, *SAV, SAVe:SETUp
*SAV (No Query Form)
(Save) Stores the state of the oscilloscope into a specified nonvola-
tile memory location. You can later use the *RCL command to
restore the oscilloscope to this saved state. This is equivalent to
selecting the Save Setup option in the Save/Recall Setup menu.
Save and Recall
*SAV <NR1>
<NR1> is an integer value in the range from 1 to 5 (TDS200 series) or
from 1 to 10 (TDS1000 and TDS2000 series) and specifies a memory
location. Any settings that have been stored previously at this
location are overwritten.
*SAV 2
saves the current settings in memory location 2.
FACtory, *RCL, RECAll:SETUp, SAVe:SETUp
SAVe:SETUp (No Query Form)
Saves the current state of the oscilloscope into the specified
nonvolatile memory location. This is equivalent to selecting the Save
Setup option in the Save/Recall Setup menu.
Save and Recall
Related Commands
Group
Syntax
Arguments
Examples
Related Commands
Group