Command Descriptions
TDS200/1000/2000 Series Oscilloscope Programmer Manual
2−139
TRIGger:MAIn:EDGE? (Query Only)
Returns the trigger coupling, source, and slope settings for the edge
trigger.
Trigger
TRIGger:MAIn:EDGE?
Trigger coupling, source, and slope settings for the main edge trigger
TRIGger:MAIn:EDGE?
might return SOURCE CH1;COUPLING DC;SLOPE RISE
TRIGger:MAIn:EDGE:COUPling
Sets or queries the type of coupling for the edge trigger. This is
equivalent to setting the Coupling option in the Trigger menu.
Trigger
TRIGger:MAIn:EDGE:COUPling { AC | DC | HFRej | LFRej |
NOISErej }
TRIGger:MAIn:EDGE:COUPling?
AC selects AC trigger coupling.
DC selects DC coupling.
HFRej coupling removes the high-frequency components of the DC
signal.
Group
Syntax
Returns
Examples
Group
Syntax
Arguments