EasyManuals Logo

Tektronix TDS1012 Programmer's Manual

Tektronix TDS1012
250 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #157 background imageLoading...
Page #157 background image
Command Descriptions
TDS200/1000/2000 Series Oscilloscope Programmer Manual
2−139
TRIGger:MAIn:EDGE? (Query Only)
Returns the trigger coupling, source, and slope settings for the edge
trigger.
Trigger
TRIGger:MAIn:EDGE?
Trigger coupling, source, and slope settings for the main edge trigger
TRIGger:MAIn:EDGE?
might return SOURCE CH1;COUPLING DC;SLOPE RISE
TRIGger:MAIn:EDGE:COUPling
Sets or queries the type of coupling for the edge trigger. This is
equivalent to setting the Coupling option in the Trigger menu.
Trigger
TRIGger:MAIn:EDGE:COUPling { AC | DC | HFRej | LFRej |
NOISErej }
TRIGger:MAIn:EDGE:COUPling?
AC selects AC trigger coupling.
DC selects DC coupling.
HFRej coupling removes the high-frequency components of the DC
signal.
Group
Syntax
Returns
Examples
Group
Syntax
Arguments

Table of Contents

Other manuals for Tektronix TDS1012

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS1012 and is the answer not in the manual?

Tektronix TDS1012 Specifications

General IconGeneral
BrandTektronix
ModelTDS1012
CategoryTest Equipment
LanguageEnglish

Related product manuals