Command Descriptions
TDS200/1000/2000 Series Oscilloscope Programmer Manual
2−87
FACtory
Setting the oscilloscope to factory default has the following impact
on the programming interface:
H Clears the Event Status Enable Register
H Clears the Service Request Enable Register
H Sets the Device Event Status Enable Register to 255
H Sets the Power On Status Clear Flag to TRUE
H Enables all Command Headers (HEADer ON)
H Sets the macro defined by *DDT to a zero-length field
H Clears the pending operation flag and associated operations
H Performs the equivalent of DATA INIT
The FACtory command does not alter the following items:
H State of the RS-232 or GPIB interface
H Selected GPIB address
H Front panel LOCK state
H Verbose state
H Display Contrast, TDS1000 and TDS2000 series only
H Calibration data that affects device specifications
H Stored settings
H Stored waveforms or data
H Hard copy parameters
H Display messages language selection
Refer to Appendix B: Factory Setup.
*PSC, *RCL, RECAll:SETUp, *RST, *SAV, SAVe:SETUp
Syntax
Examples
Related Commands