EasyManua.ls Logo

Tektronix TDS200 Series - Page 111

Tektronix TDS200 Series
250 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Descriptions
TDS200/1000/2000 Series Oscilloscope Programmer Manual
2−93
HARDCopy:PORT?
might return RS232 as the selected hard copy output port.
HDR
This command is identical to the HEADer query and is included for
compatibility with other Tektronix oscilloscopes.
HEADer
Sets and queries the Response Header Enable State that causes the
oscilloscope to either include or omit headers on query responses.
This command does not affect IEEE Std 488.21987 Common
Commands (those starting with an asterisk); they never return
headers.
Miscellaneous
HEADer { <NR1> | OFF | ON }
HEADer?
ON or <NR1> 0 sets the Response Header Enable State to true. This
causes the oscilloscope to include headers on applicable query
responses. You can then use the query response as a command.
OFF or <NR1> = 0 sets the Response Header Enable State to false.
This causes the oscilloscope to omit headers on query responses so
that only the argument is returned.
HEADer OFF
causes the oscilloscope to omit headers from query responses.
Examples
Group
Syntax
Arguments
Examples

Table of Contents

Related product manuals