Command Descriptions
2−96
TDS200/1000/2000 Series Oscilloscope Programmer Manual
<NR3> is the position in seconds. This value is the difference between
the trigger point and the center graticule. Positive values place the
trigger before the center graticule.
HORizontal:DELay:POSition 2.0E–6
sets the window position to 2 ms before the center graticule.
HORizontal:DELay:POSition?
might return –1.0E–3, indicating that the window position is 1 ms
after the center graticule.
HORizontal:DELay:SCAle
Sets the time per division for the oscilloscope window time base.
This command is equivalent to setting SEC/DIV when Window Zone
or Window is selected from the Horizontal menu.
Horizontal
HORizontal:DELay:SCAle <NR3>
HORizontal:DELay:SCAle?
<NR3> is the time per division. The range depends on the oscillo-
scope model. The acceptable values are in a 12.55 sequence. If the
window time base scale is set slower than the main time base scale,
both the main and window time base scales are set to the window
scale value.
HORizontal:DELay:SCAle 2.0E–6
sets the window scale to 2 ms per division.
Arguments
Examples
Group
Syntax
Arguments
Examples