Command Descriptions
2−162
TDS200/1000/2000 Series Oscilloscope Programmer Manual
WFMPre:ENCdg
Sets or queries the type of encoding for waveform data transferred
with the CURVe command. Changing WFMPre:ENCdg also changes
DATa:ENCdg.
Waveform
WFMPre:ENCdg { ASC | BIN }
WFMPre:ENCdg?
ASC specifies ASCII curve data.
BIN specifies binary curve data.
WFMPre:ENCdg ASC
specifies that the waveform data is in ASCII format.
WFMPre:ENCdg?
might return BIN, indicating that the waveform data is in binary
format.
DATa:ENCdg
WFMPre:NR_Pt? (Query Only)
Returns the number of points that are in the transmitted waveform
record, as specified by DATa:SOUrce. The number of points depends
on DATa:STARt, DATa:STOP, and whether DATa:SOUrce is YT or
FFT. NR_Pt is at most 2500 for YT and 1024 for FFT. NR_Pt is
always at least one.