Status and Events
TDS200/1000/2000 Series Oscilloscope Programmer Manual
3−21
Table 3−5: Execution error messages – EXE bit 4 (Cont.)
Code Message
2243 Waveform requested is not a data source
2244 Waveform requested is not turned on
2245 Saveref error, Selected channel is turned off
2246 Saveref error, Selected channel data invalid
2248 Saveref error, Source reference data invalid
2260 Calibration error
2301 Cursor error, Off screen
2302 Cursor error, Cursors are off
2303 Cursor error, Cursor source waveform is off
Table 3-6 lists the device errors that can occur during oscilloscope
operation. These errors may indicate that the oscilloscope needs
repair.
Table 3−6: Device error messages – DDE bit 3
Code Message
300 Device−specific error
310 System error
311 Memory error
313 Calibration memory lost
314 Save/recall memory lost
315 Configuration memory lost
350 Queue overflow (does not set DDE bit)
361 Parity error in program message (check parity)