Command Descriptions
TDS200/1000/2000 Series Oscilloscope Programmer Manual
2−55
CH<x>:PRObe?
1 sets the specified oscilloscope channel to 1X attenuation.
10 sets the specified oscilloscope channel to 10X attenuation.
100 sets the specified oscilloscope channel to 100X attenuation.
1000 sets the specified oscilloscope channel to 1000X attenuation.
<NR1>
CH2:PRObe 1000
sets channel 2 to x1000 attenuation.
CH1:PRObe?
might return CH1:PROBE 10
CH<x>:SCAle
Sets or queries the vertical gain of the specified oscilloscope
channel. This command is equivalent to adjusting the front-panel
VOLTS/DIV knob.
Vertical
CH<x>:SCAle <NR3>
CH<x>:SCAle?
<NR3> is the gain, in volts per division. The range is 5 V/div to
2 mV/div when using a 1X probe.
Arguments
Returns
Examples
Group
Syntax
Arguments