Command Descriptions
2−64
TDS200/1000/2000 Series Oscilloscope Programmer Manual
CURSor:VBArs:POSITION<x>?
<x> specifies which cursor to position. Correct values are 1 and 2.
<NR3> specifies the cursor position in the units specified by the
CURSor:VBArs:UNIts command. The position is relative to the
trigger except when the cursor source is a math FFT waveform.
The cursor position is limited to the graticule whenever an attempt is
made to move it outside the graticule.
CURSor:VBArs:POSITION2 9.00E–6
positions the second vertical bar cursor at 9 ms.
CURSor:VBArs:POSITION1?
might return 1.00E–6, indicating the first vertical bar cursor is at
1 ms.
CURSor:VBArs:UNIts
Sets or queries the units for the vertical bar cursors.
NOTE. When Trigger View is active, CURorS:VBArs:UNIts?
generates event 221(Settings conflict).
Cursor
CURSor:VBArs:UNIts { SECOnds | HERtz }
CURSor:VBArs:UNIts?
SECOnds specifies units of time.
Arguments
Examples
Group
Syntax
Arguments