EasyManua.ls Logo

Tektronix THM565 - Page 40

Tektronix THM565
382 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DCM330 Performance Verification
8
Handheld and Benchtop Instruments Basic Service
To check the frequency accuracy, perform the following steps.
1. Set the meter function to Hz.
2. Select the appropriate coil as necessary to multiply the AC Current calibrator
output to 20 A.
3. Position the clamp around the current loop of the AC current calibrator and
release the clamp trigger. Ensure that the clamp is entirely closed.
4. Verify that the display reads within the specified Low and High Limits for
each of the frequencies listed in Table 9 on page 10 and record the reading.
5. Disconnect the calibrator.
Frequency Check

Table of Contents

Other manuals for Tektronix THM565

Related product manuals