DMM800 Series Performance Verification
Handheld and Benchtop Instruments Basic Service
9
Test Equipment
The performance verification procedures use external traceable test equipment to
directly check warranted characteristics. If you substitute equipment, always
choose instruments that meet or exceed the minimum requirements listed in
Table 8.
Alternative test equipment must meet or exceed the intended minimum
requirements. If you substitute equipment, you may need to modify the
performance verification procedures.
NOTE. Before beginning the performance verification procedures, warm up the
test equipment according to the manufacturer’s recommendations.
Table 8: Test Equipment
Description Minimum Requirements Example Product
ÁÁÁÁÁÁ
ÁÁÁÁÁÁ
Universal Calibration
System
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
Resolution & accuracy 4 times
greater than the multimeter
display reading.
ÁÁÁÁÁÁÁ
ÁÁÁÁÁÁÁ
Wavetek 9100 with 9105 lead
set.
AC and DC volts measurement
1
AC and DC current measurement
Resistance measurement
1
Capacitance measurement
Sinewave generation
Squarewave generation
Thermocouple adapter K Type Tektronix ATK01
Capacitance Standard Optional
1
Choose 4-wire measurement setup if available.