DCM910 Specifications
4
Handheld and Benchtop Instruments Basic Service
Table 3: Certifications and Compliances
Certifications
anadian Standards Association certified to Standard
SA 1
1
.1, Standard
UL3111-1 for Electrical and Electronic Measuring and Testing Equipment, and
IEC1010-2-032 particular requirements for hand-held current clamps for electrical
measurement and test.
Overvoltage Category Category: Examples of Products in this Category:
CAT III Distribution-level mains, fixed installation
CAT II Local-level mains, appliances, portable equipment
CAT I Signal levels in special equipment or parts of equipment,
telecommunications, electronics
Pollution Degree 2 Do not operate in environments where conductive pollutants may be present.
Table 4: Environmental Characteristics
Characteristic Description
Temperature
Operating
0° to 50° C (32° to 122° F), <75% relative humidity
Nonoperating
–20° C to +60° C (21° to 140° F), <80% relative humidity
Temperature Coefficient
0.2 × (specified accuracy) per °C at <18° C or >28° C
Maximum Altitude (Operating) 2,000 m (6,562 ft)