EasyManua.ls Logo

Tektronix WVR6020 - Page 112

Tektronix WVR6020
138 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Video and General Performance Verication
SD Jitter Readout
Accuracy (Opt
ions EYE or
PHY)
This test check
s the SD jitter readout accuracy.
1. Connect the DVG1 Option S1 output (typically the second output) to the SDI
A input and ter
minate the loopthrough connector.
2. Set the DVG1 to generate a 525-270 100% Color Bars signal.
3. Press the left a rrow key and then select Display Type > Eye Display.
4. Press the left arrow key and then select Jitter1 HP Filter > 100 Hz.
5. If the jitter readout is not displayed, press the left arrow key, and then select
Jitter Meter > Meter & Readout.
6. Adjust the DVG1 SDI module parameters as follows:
a. Set the Jitter Freq to 9950 Hz.
b. Set the J
itter Ampl to 11.1 ns (3.00 UI).
7. Check that the jitter readout is between 2.65 UI and 3.35 UI.
8. Record Pass or fail in the test record for SD Jitter Readout Accuracy.
9. Reset the generator.
98 Waveform Rasterizers Specications and Performance Verication

Table of Contents

Other manuals for Tektronix WVR6020

Related product manuals