Video and General Performance Verication
Composite Analog Field
Rate Tilt (Option CPS)
Apply a eld rat
e square wave and use cursors to measure the tilt.
1. Press FACTORY.
2. Press CONFIG, select Graticules and set Cmpst Waveform Graticule Units to
mV.
3. Connect the precision video generator to one of the Composite A input
connectors and terminate the other connector with the precision terminator.
4. Set the precision generator for Field Square Wave.
5. Press the C MPST A button, then the CURSOR button, and then the FULL
button.
6. Adjust cursors to be near the top of the waveform.
7. Use the vertical position knob to move the top of the waveform to the zero
graticule.
8. Press the 5x gain button and the FIELD button.
9. Use cursors to measure the tilt in the waveform.
10. Recor
d the value in the test record.
Composite Analog Line
Rate tilt (Option CPS)
Apply a eld rate square wave and use cursors to measure the tilt.
1. Press FACTORY.
2. Press CONFIG, select Graticules, and then set Cmpst Waveform Graticule
Units to mV.
3. Connect the precision video generator to one of the Composite A input
connectors and terminate the other connector with the precision terminator.
4. Set the precision generator for 100% Flat Field.
5. Press the CMPST A, CURSOR, and FULL buttons.
6
.
A
djust the cursors to be near the top of the waveform.
7. Use the VERTICAL position knob to move the top of the waveform to the
zero graticule.
8. Press 5X.
9. Use the cursors to measure the tilt in the waveform, excluding the rst and
last 5 µs of the bar.
10. Record the value in the test record.
104 Waveform Rasterizers Specications and Performance Verication