Telstar Bio II Advance Plus - User Manual

Telstar Bio II Advance Plus
244 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
USER MANUAL
MANUAL DE USUARIO
MANUEL D'UTILISATION
BEDIENUNGSANLEITUNG
Microbiological Safety Cabinets
Cabinas de Seguridad Microbiológica
Cabines de sécurité microbiologique
Mikrobiologische Sicherheitswerkbänke
Bio II Advance Plus

Table of Contents

Questions and Answers

  • L
    Lisa EllisAug 14, 2025
    What to do if the Telstar Laboratory Equipment display does not turn on?
    • P
      Pamela CohenAug 14, 2025
      First, ensure the power cord is securely connected to both the cabinet and the wall socket. Next, verify that the wall socket has power. If these steps don't resolve the issue, a specialized technician should replace the main control.
  • J
    Jamie RiceAug 17, 2025
    What to do if the Telstar Laboratory Equipment shows insufficient exhaust flow alarm LOW INFLOW?
    • R
      Robert WhiteAug 17, 2025
      If you're seeing a 'LOW INFLOW' alarm, start by visually inspecting the exhaust for any obstructions. Then, check the front perforation in the working area, ensuring it's free from blockage. Also, consider the distance between the cabinet exhaust and the lab ceiling; increasing this space may help. Check the probe position and make sure it is correct.
  • L
    Lori FlemingAug 21, 2025
    What to do if the Telstar Laboratory Equipment shows insufficient laminar down flow alarm LOW LAM. FLOW?
    • J
      Jody BeckerAug 21, 2025
      If you're getting a 'LOW LAM. FLOW' alarm, start by visually inspecting the front and rear perforation in the work zone for any obstructions. Also, check the probe position to be sure it is correct and remove any obstruction to the probe.
  • S
    sherry31Aug 25, 2025
    What to do if the Telstar Laboratory Equipment shows excessive laminar down flow alarm HIGH LAM. FLOW?
    • M
      Michael FrancisAug 26, 2025
      If the alarm 'HIGH LAM. FLOW' is on, check the probe position to be sure it is correct. Also, check and remove any airflow in front of the cabinet that distorts the reading of the sensor.
  • D
    Debra BartonAug 28, 2025
    What to do if the Telstar Bio II Advance Plus Laboratory Equipment glass is out of position and shows alarm WORK POSITION?
    • A
      Allen GoodwinAug 28, 2025
      If the alarm 'WORK POSITION' is on, assemble the glass in working position.
  • J
    Jonathon ClarkSep 1, 2025
    What to do if the Telstar Laboratory Equipment open frontal shows alarm WORK POSITION?
    • C
      Christopher TaylorSep 1, 2025
      If the alarm 'WORK POSITION' is on, slide down the glass to closed position and check that the frontal contacts are not loose and remove any type of obstruction.
  • C
    Courtney SmithSep 5, 2025
    What to do if the Telstar Bio II Advance Plus Laboratory Equipment light does not turn on?
    • T
      tayloraprilSep 5, 2025
      Check that the lamp is properly connected and change the led bar.
  • M
    Miguel DavidsonSep 9, 2025
    What to do if the Telstar Bio II Advance Plus UV germicidal lamp does not turn on (optional accessory)?
    • H
      Hayley JonesSep 9, 2025
      Check that the front slides properly. Also, check that the lamp holder is connected properly and change the UV lamp. Change the reactance.

Summary

Description of the Safety Cabinets

Installing Instructions

Placing on Site

Instructions on proper placement and positioning of the cabinet.

Electrical Connection and Services

Details on power requirements and connection procedures.

Installation Tests

Mandatory tests to ensure correct operation after installation.

Operating Instructions

Start-Up

Step-by-step guide for powering on and initiating the cabinet.

Control Keyboard-Display

Overview of the control panel interface, functions, and displays.

Maintenance

Absolute Filters

Details on HEPA filter lifespan, replacement, and handling.

Cleaning and Disinfection

Surface Cleaning and Disinfection

Procedures for routine cleaning of cabinet surfaces.

Tests

Filter Integrity Tests

Verifies the integrity of both downflow and exhaust HEPA filters.

Air Velocity Tests

Measures downflow and inflow air velocities and tests alarm set points.

Troubleshooting

Telstar Bio II Advance Plus Specifications

General IconGeneral
TypeBiological Safety Cabinet
Filter TypeHEPA H14
Front Opening200 mm
UV LightOptional
CertificationEN 12469
Work Area Dimensions900 mm width
Noise Level< 65 dB(A)
Power Supply50Hz