Telstar Bio II Advance - User Manual

Telstar Bio II Advance
229 pages
Go to English
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
MANUAL DE USUARIO
USER MANUAL
MANUEL D'UTILISATION
BEDIENUNGSANLEITUNG
Cabinas de Seguridad Microbiológica
Microbiological Safety Cabinets
Cabines de sécurité microbiologique
Mikrobiologische Sicherheitswerkbänke
Bio II Advance

Table of Contents

Questions and Answers

  • T
    Thomas ScottAug 17, 2025
    How to fix the “LOW EXH. FLOW” alarm on my Telstar Bio II Advance Laboratory Equipment?
    • E
      Elizabeth GarzaAug 17, 2025
      If the exhaust flow alarm “LOW EXH. FLOW” is triggered on your Telstar Laboratory Equipment, it could be due to several reasons. First, visually inspect the exhaust for any obstructions. Also, check the front perforation in the working area for any blockages. The distance between the cabinet exhaust and the lab ceiling might be too small; consider increasing this space, modifying the false ceiling, or installing a lower support table. The exhaust filter could be clogged, so check its condition and the date of the last filter change. Ensure the probe is correctly positioned. If you have a ducted cabinet, the ducting might be too long; in this case, consider fitting an additional fan at the end of the duct.
  • A
    Amanda JacksonAug 20, 2025
    What to do if the “LOW LAM. FLOW” alarm appears on Telstar Laboratory Equipment?
    • B
      Brian ReyesAug 20, 2025
      If the laminar down flow alarm “LOW LAM. FLOW” is activated on your Telstar Laboratory Equipment, start by visually inspecting the front and rear perforation in the work zone for any obstructions. Next, check the laminar down flow filter for clogging. Also, verify that the probe is correctly positioned (sensor parallel to the flow) and free of any obstructions.
  • W
    William ParksAug 22, 2025
    Why is the “LOW EXH. FLOW” alarm on and the exhaust not working on my Telstar Laboratory Equipment?
    • M
      michelleespinozaAug 22, 2025
      If the exhaust is not working properly and the alarm “LOW EXH. FLOW” is showing on your Telstar Laboratory Equipment, first visually check the exhaust for any obstructions. Also, ensure the probe is correctly positioned and connected. If the exhaust fan has stopped, this could also be the cause.
  • L
    Lisa WardAug 25, 2025
    What causes the “LOW LAM. FLOW” alarm and laminar down flow malfunction in Telstar Bio II Advance Laboratory Equipment?
    • D
      Deborah PhillipsAug 26, 2025
      If the laminar down flow is not working properly and the alarm “LOW LAM. FLOW” is triggered on your Telstar Laboratory Equipment, verify the probe's position, ensuring it's correctly aligned and free from obstructions. Additionally, check if the laminar down flow fan has stopped.
  • N
    Nathan FosterAug 27, 2025
    How do I resolve the “HIGH LAM. FLOW” alarm on my Telstar Laboratory Equipment?
    • D
      Donna LopezAug 28, 2025
      If the laminar down flow alarm “HIGH LAM. FLOW” is triggered on your Telstar Laboratory Equipment, check the probe position to ensure it is correct. Also, check for and eliminate any air currents in front of the cabinet that might be distorting the sensor reading.
  • M
    Mary LopezAug 30, 2025
    Why doesn’t the light turn on in my Telstar Bio II Advance Laboratory Equipment?
    • K
      Kevin WoodardAug 30, 2025
      If the light on your Telstar Laboratory Equipment does not turn on, ensure the connection in the control card is correct. If that's not the problem, slightly turn the fluorescent light to ensure good contact or change the fluorescent bulb. If these steps don't work, the ballast might be damaged and needs to be changed.
  • N
    Nicole HumphreySep 1, 2025
    What to do if the UV lamp doesn’t turn on in my Telstar Bio II Advance Laboratory Equipment?
    • E
      Edward FrancoSep 1, 2025
      If the UV germicidal lamp does not turn on in your Telstar Laboratory Equipment (if it's an optional accessory), check that the front slides properly into the closed position. If the glass is in the correct position, check the lamp holder is connected properly. If it still doesn’t work, the UV lamp may need to be changed, or the ballast might be damaged.
  • H
    henrycolleenSep 4, 2025
    Why won't the front of my Telstar Laboratory Equipment move?
    • S
      sreedSep 4, 2025
      If your Telstar Laboratory Equipment's front does not move, visually check for any obstructions that may be blocking its movement.
  • D
    Devin HortonSep 6, 2025
    What should I do if the display on my Telstar Bio II Advance Laboratory Equipment doesn't turn on?
    • B
      butlersarahSep 6, 2025
      If the display on your Telstar Laboratory Equipment does not turn on, check that the power cord is properly connected to both the cabinet and the wall socket. Also, verify that there is power at the socket.
  • D
    david26Sep 8, 2025
    How do I fix the “WORK POSITION” alarm on my Telstar Bio II Advance Laboratory Equipment when the glass is out of position?
    • S
      Stacey WellsSep 8, 2025
      If the glass is out of position and the alarm “WORK POSITION” is triggered on your Telstar Laboratory Equipment, assemble the glass in the working position.

Summary

TECHNICAL FEATURES

INSTALLING INSTRUCTIONS

PRECAUTIONS FOR FIRST CONNECTION

ELECTRICAL CONNECTION AND SERVICES

CABINET EXHAUST DUCTING

OUTSIDE DUCTING VIA A CANOPY

INSTALLATION TESTS

OPERATING INSTRUCTIONS

START-UP

Step-by-step guide for turning the safety cabinet on.

GENERAL WORKING RECOMMENDATIONS

ERGONOMICS

DEFINITION OF THE WORKING AREA

CONTROL KEYBOARD-DISPLAY

DISCONNECTION

MAINTENANCE

MAINTENANCE TABLE

Schedule and frequency for routine maintenance tasks.

CLEANING AND DISINFECTION

DISINFECTION WITH FORMALDEHYDE

Detailed procedure for using formaldehyde for cabinet decontamination.

TESTS

DOWNFLOW FILTERS INTEGRITY TEST

Procedure to verify the integrity of downflow HEPA filters.

EXHAUST FILTERS INTEGRITY TEST

Procedure to verify the integrity of exhaust HEPA filters.

DOWNFLOW AIR VELOCITY TEST

Measures and verifies the downflow air velocity.

SET POINT ALARM DOWNFLOW AIR VELOCITY TEST

Tests the alarms for downflow air velocity set points.

INFLOW AIR VELOCITY TEST

Measures and verifies the inflow air velocity.

TROUBLESHOOTING

Telstar Bio II Advance Specifications

General IconGeneral
BrandTelstar
ModelBio II Advance
CategoryLaboratory Equipment
LanguageEnglish