How to fix the “LOW EXH. FLOW” alarm on my Telstar Bio II Advance Laboratory Equipment?
- EElizabeth GarzaAug 17, 2025
If the exhaust flow alarm “LOW EXH. FLOW” is triggered on your Telstar Laboratory Equipment, it could be due to several reasons. First, visually inspect the exhaust for any obstructions. Also, check the front perforation in the working area for any blockages. The distance between the cabinet exhaust and the lab ceiling might be too small; consider increasing this space, modifying the false ceiling, or installing a lower support table. The exhaust filter could be clogged, so check its condition and the date of the last filter change. Ensure the probe is correctly positioned. If you have a ducted cabinet, the ducting might be too long; in this case, consider fitting an additional fan at the end of the duct.