USER & SERVICE MANUAL
MANUAL DE USUARIO Y MANTENIMIENTO
L
aboratory Freeze Dryers
Liofilizadores de laboratorio
LyoQuest
Ly
oQuest Arctic
DOMINIQUE DUTSCHER SAS

Table of Contents

Questions and Answers

  • N
    Natalie BallAug 10, 2025
    What to do if my Telstar LyoQuest -85 Plus pump does not reach the vacuum pressure limit?
    • R
      Roy FigueroaAug 11, 2025
      If your Telstar Laboratory Equipment vacuum pump isn't reaching the vacuum pressure limit, it might be due to a lack of oil. Check the oil level and add more until it reaches the appropriate quantity. Alternatively, the oil could be contaminated. In this case, empty the old oil and refill the pump with new oil.
  • J
    jlopezAug 17, 2025
    Why my Telstar LyoQuest -85 Plus vacuum pump does not start?
    • R
      rebeccaperryAug 17, 2025
      If the Telstar Laboratory Equipment vacuum pump does not start, several issues could be the cause. First, check the connection of the on/off switch located on the motor of the pump. Also, examine the electrical connection terminals of the pump and the corresponding circuit-breaker. The environmental temperature might be too high (over 35ºC), which could cause the circuit to break. Finally, the pump's starter protector may be blocked because the condenser isn't at the correct temperature. The minimum temperature to switch on the pump is: LQ-55
  • L
    Lindsey BrownAug 25, 2025
    How to prevent abnormal heating of Telstar Laboratory Equipment pump?
    • B
      Brian FergusonAug 26, 2025
      If the Telstar Laboratory Equipment pump is heating abnormally, it may be due to poor ventilation or a high environmental temperature (above 35ºC). Try placing the pump in an area with better ventilation to resolve this issue.
  • C
    Carol KnightSep 2, 2025
    What to do if I see mist leak from the emission channel of Telstar LyoQuest -85 Plus?
    • J
      Jennifer McgeeSep 2, 2025
      If you notice mist leaking from the emission channel of your Telstar Laboratory Equipment, it could be because the Gas-Ballast is opened too much; try closing it. Another possible reason is a leak somewhere in the system; try to locate the leak.

Summary

Description of the Freeze Dryer

Technical features

Presents detailed technical specifications for vacuum pump, ice condenser, and refrigerating unit.

Installation Instructions

Electrical connection

Details on ensuring correct power supply and connecting the freeze-dryer to the mains.

Working Principle

Selectable processes

Details the Freezing, Cool+Vacuum, and Heat plates processes with their parameters.

Operation and Control Instructions

Start up

Step-by-step guide for powering on the equipment and accessing the main screen.

Freeze-drying

Details the Freeze-drying window, its components, and control buttons for automatic and semiautomatic cycles.

Semiautomatic cycle

Guides the user through selecting processes and viewing set points in the Semiautomatic cycle window.

Alarms

Details the Alarms window for viewing active, muted, and corrected alarms with status and text.

Settings

Allows users to adjust equipment and display settings like vacuum pump time, temperature, and manifold options.

Recipe edition

Enables creating and editing lyophilization recipes, managing up to 10 steps per recipe.

Load recipe

Allows selecting and loading a previously created recipe for automatic execution.

Users

Manages user accounts, allowing login, password changes, and user creation/editing.

Users management

Allows administrators to add, edit, or remove user accounts.

Maintenance

Cleaning and disinfection of the unit

Provides detailed procedures for cleaning and disinfecting the freeze dryer and its components.

Vacuum unit maintenance

Covers oil checks, oil condition, level inspection, leak checks, and pump body cleaning.

Maintenance table

Provides a table with recommended periodicities for vacuum unit, refrigerating unit, and freeze dryer maintenance tasks.

Troubleshooting

Vacuum unit

Troubleshooting guide for common vacuum pump issues like failure to start or obtain vacuum.

Refrigerating unit

Troubleshooting for refrigerating unit problems, including startup issues and compressor stoppages.

Freeze-dryer

Troubleshooting for issues like failing to reach working vacuum or gauge errors.