J
Jeanette BeardJul 27, 2025
Why is the warning triangle flashing on my TensCare Personal Care Product and I cannot increase the intensity?
- CcolejeffreyJul 27, 2025
If the warning triangle is flashing and you can't increase the intensity on your TensCare Personal Care Product, especially when using EMS Manual Programme 10 or 11, it might be because the 'Ramp Up Time', 'Work Time', and 'Ramp Down Time' do not equal 'Rest Time' in ALTERNATING mode. Also, verify the electrode pad positions. If the pads are safely positioned, you can press the SKIP key to override the warning.