    Matthew MurphyAug 10, 2025
    What to do if TensCare Personal Care Products shows no sensation and LEADS alarm?
    • T
      Trevor BlackwellAug 10, 2025
      If you experience no sensation and the LEADS alarm is showing, it could be due to a break in one of the lead wires. Try testing the unit by dampening your hand with water and a little table salt, then firmly squeeze the probe ensuring your skin covers the metal parts. Carefully increase the strength until you feel something. Alternatively, the skin may be too dry, causing poor conductivity. In this case, use a water-based lubricant like TensCare Go Gel to improve conduction, or cross your legs and squeeze to increase pressure on the probe to improve the connection.
  • B
    Brenda MillerAug 15, 2025
    How to fix TensCare Personal Care Products controls that won’t work?
    • D
      daniellewoodAug 16, 2025
      If the controls on your TensCare Personal Care Products device aren’t working, it might be because the keypad is locked. Check the display for the 'LOCK' indicator. If it's visible, press and hold the button. If 'LOCK' isn't showing, try removing and then replacing the batteries.
  • D
    David BrownAug 20, 2025
    How to fix TensCare Unicare Personal Care Products when there is no display?
    • N
      nhamiltonAug 20, 2025
      If your TensCare Personal Care Products device has no display, it could be due to flat batteries. Try recharging the batteries. Alternatively, the battery may be damaged, in which case you should contact the supplier.
  • S
    Sonya LewisAug 26, 2025
    What to do if TensCare Unicare has no sensation and no LEADS alarm showing?
    • N
      Nicholas CarterAug 26, 2025
      If you are experiencing no sensation and no LEADS alarm is showing on your TensCare Personal Care Products device, ensure that the intensity level is high enough. Most people start to feel the stimulation in their hand around 25 mA, and with the probe inserted, you may need to increase the intensity to around 40 mA – 60 mA.
  • S
    Samantha LinAug 30, 2025
    Why TensCare Unicare shows low battery display?
    • C
      Cynthia FosterAug 30, 2025
      If your TensCare Personal Care Products device is showing a low battery display, it is likely due to low batteries. Recharge the batteries.
  • E
    ecastilloSep 4, 2025
    How to fix TensCare Personal Care Products when there is no sensation on one side of the probe?
    • G
      gouldjenniferSep 4, 2025
      If you're experiencing no sensation on one side of the probe (or electrode) with your TensCare Personal Care Products device, the probe position may not be optimal. Try slightly adjusting the probe's position or exchanging the connection of the wires in the probe.

TensCare Unicare Specifications

General IconGeneral
TechnologyEMS (Electrical Muscle Stimulation)
Power SourceBattery
ChannelsSingle Channel
Intensity LevelsAdjustable
WeightApprox. 100g
Intended UsePelvic floor muscle stimulator