EasyManua.ls Logo

TestEquity FS Series - Page 15

Default Icon
19 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 4 Maintenance
TestEquity FS Series Ovens Page 4-3
TEMPERATURE (continued)
Indicated chamber temperature
unstable
1. ±0.1°C variation is normal.
2. Make sure fan is working. Verify movement of air in chamber.
3. Room temperature is radically changing either door opening or
room airflow from heaters or air conditioning. Stabilize the ambient
conditions.
4. This may happen if exhaust stack is 100% open or if power exhaust
is cycling adjust stack to at least ¼ closed.
5. Sensor miss-located, damaged or wires may be damaged. Check
mounts for control and overtemperature sensors, then trace wires or
tubing between sensors and controls.
6. Defective temperature controller. Call Customer Service.
7. OVERTEMP PROTECTION set too low. Be sure that
OVERTEMP PROTECTION is set more than 5 degrees over
temperature set point; check if OVERTEMPERATURE light is on
continuously; turn OVERTEMP PROTECTION knob completely
clockwise to see if problem solved then follow operating
instructions for correct setting.
8. Electrical noise. Remove nearby sources of RFI including motors,
arcing relays or radio transmitters
9. Bad connection on temperature sensor or faulty sensor. Check
connectors for continuity and mechanical soundness while watching
display for erratic behavior; check sensor and wiring for mechanical
damage.
10. Bad connections or faulty solid state relay. Check connectors for
mechanical soundness and look for corrosion around terminals or
signs of arcing or other visible deterioration.
MECHANICAL
Motor doesn’t move
1. If shaft spins freely: check connections to motor and check voltage
to motor.
2. If shaft rubs or is frozen, relieve binding and retest.
Motor makes noise
1. If noise is from the motor, tap the top of motor shaft with ball peen
hammer.
2. If the sound gets worse, tap the other end of the shaft avoiding
touching the blower wheel.
3. If there is no change, call Customer Service.
Door not sealing
1. Adjust hinge blocks, chamber latch bracket or twist the door.
2. Confirm that unit has not been damaged and body is not square.
3. Check physical condition of gasket for tears or punctures.