9. Technical data16
9. Technical data
Instrument
Characteristic Value
Parameters Temperature (°C/°F)
Measuring range Pt100 probe: -50...+300°C / -58...+572°F
NTC probe: -50...+120°C / -58...+248°F
Resolution 0.1°C / 0.1°F
Accuracy º See SYSTEM ACCURACY
Probe 1x mini DIN socket for Pt100 or NTC temperature probe
Measuring rate 2/s
Operating temperature range -20...+50°C / -4...+122°F
Storage temperature -30...+70°C / -22...+158°F
Voltage supply 1x 9V monobloc battery/rech. battery
Battery life approx. 70h
Protection class with TopSafe (accessory part) and probe connected: IP65
EC Directive 89/336/EEC
Warranty 2 years
System accuracy
Measuring range Instrument Probe System
Measuring instrument + NTC temperature probe
-50.0°C...-25.1°C ±1% of reading ±0.7% of reading ±1.8% of reading
-25.0...+40.0°C ±0.2°C ±0.2°C ±0.5°C
+40.1...+80.00°C ±0.3°C ±0.4°C ±0.8°C
+80.1...+120.0°C ±0.5°C ±0.6°C ±1.2°C
Measuring instrument + Pt100 temperature probe
-50.0...-25.1°C ±0.2°C ±0.3°C ±0,6°C
-25.0...+40.0°C ±0.2°C ±0.2°C ±0.5°C
+40.1...+140.0°C ±0.2°C ±0.4°C ±0.7°C
+140.1...+200.0°C ±0.2°C ±0.6°C ±0.9°C
+200.1...+300.0°C ±0.3°C ±0.8°C ±1.2°C