Multi-purpose clip
Holder for positioning
stick in duct
Ø 12 mm
The humidity sensor is
protected by a quick
twist of the bottom of
the probe stem
the
protective cap is closed.
Should only be left open
for the duration of the
measurement.
Swivel head. The dis-
play can always be
read.
Commissioning
Remove the protective film on the display.
Switching on
Press button:
Segmenttest
Current reading
Changing the parameters
Press button repeatedly:
Temperature Dew point Humidity
Switching off
Keep button pressed for 3 seconds.
Selecting between °C/°F
When switching on keep button pressed until the
segment test is finished, instrument switches to
configuration menue. Press button again the
instrument alternates between °C and °F. Wait
2 sec. before pressing the button again. The
selected unit then applies.
Technical data
Type of application: ..................................................Short measurement
Measuring range: ................................................5 to 95 %RH / 0 to +50°C /
-20 to +50 °C td
+32 to +122°F/-4 to +122°F td
Resolution: ........................................................0.1 %RH / 0.1°C/ 0.1°F
Accuracy: ......................................................(System accuracy at 25°C/
+77 °F ) ±3 %RH / ±0.5°C/±0.9°F
Parameters: ..........................................................................%RH, °C, °F
Ambient temperature: ....................................0 to +50 °C/32 to +122°F
Storage temperature: ..................................-20 to +70°C/-4 to +160 °F
Battery type:................................................................................3 x AAA
Battery lifetime: ..............................................................Approx. 1000 h
Probe stem:................................................................................Ø 0.5 in
Length: ........................................................................................4.9 in
Warranty: ....................................................................................2 years
CE guideline......................................................................2004/108/EEC
Please read before using instrument
• Observe measurement ranges of sensor
• Do not exceed maximum storage and operating
temperatures (e.g. protect measuring instrument
from direct sunlight).
• Inexpert handling cancels your warranty.
Changing the battery
The battery has to be changed if the symbol
appears in the display during the measurement.
Instrument is switched off.
1 Open battery compartment cover.
2 Remove spent batteries and insert new batteries
(3x AAA) into the battery compartment. Observe
polarity!
3 Close battery compartment cover.
Auto Off function
If no button has been pressed in approx. 10
minutes, the instrument switches itself off
automatically.
Instruction manual en