10 Technical data for testo 625
33
10 Technical data for testo 625
Please do not use the humidity probes in condensing atmospheres. For
continuous use in high-humidity ranges
> 80% RH at ≤ 30°C for > 12 h
> 60% RH at > 30°C for > 12 h
please get in touch with Testo Service or contact us via the Testo
website.
Feature Value
Measurement parameters
%RH
°C, °F
°C wb, °C dp
Accuracy
±2.5 %RH (5 to 95 %RH)
±0.5°C
Resolution
0.1 %RH
0.1 °C
Measuring range
0 to 100.0 %RH
-20 to 60 °C
Operating temperature
-20 to +50 °C (instrument)
-20 to +60 °C (probe)
Storage temperature
-20 to +50 °C
Operating humidity
0 to 80 %RH / For indoor use only
IP class
IP20
Level of contamination
PD2
Max. operating altitude
2000 m above sea level
Nominal output
2 W @ 4,5 V DC
Battery type
3 x 1.5 V AA battery
(included in the scope of delivery)
Battery life
>100 h
Dimensions
208 x 60 x 28 mm
Probe 12 x 15 mm
Weight
199 g