6 Using the product
36
Button Function Description
Sub-level zoom Sub-level zoom with additional
functions is displayed.
: menu, see above.
: zoom in, see above.
: zoom out, see above.
: display buttons for moving the
zoomed image. To move, touch the
buttons shown in the image.
: close sub-level zoom.
Delete image
Function only available when the
image is saved.
Pos: 70 /TD/Übersc hriften/6.2 Menü-Funk tionen @ 0\mod_118639 0880578_79.doc x @ 2125 @ 2 @ 1
6.2. Menu functions
Pos: 71 /TD/Produk t verwenden/WBKs/ testo 885/885, 890: Messfunk tionen @ 9\mod_1305534 550459_79.docx @ 80536 @ 355 5555555555 @ 1
6.2.1. Measurement
Measurement range
The measurement range can be set to adapt to the respective area
of application. If measurement range 3 is selected (high-
temperature measurement range), the aperture in the lens is
automatically reduced to protect the detector against damage.
CAUTION
Detector may be damaged!
> If measurement range 1 and 2 are set: do not point the imager
at objects with temperatures above 550°C (1022°F).
> If measurement range 3 is set: do not point the imager at
objects with temperatures above 1400°C (2552°F).