7 Using the product
14
Setting the unit
Configuration mode is open, “UNITS” is displayed.
Press ◄ / ► to choose between metric (“METR”) and imperial
(“IMPER”) units of measurement and confirm with MENU/ENTER.
The currently set unit flashes.
Press ◄ / ► to set the required unit and confirm with MENU/ENTER.
Setting alarm thresholds
Alarm thresholds can only be set for the differential temperature (ΔT).
Configuration mode is open, “ min” is displayed.
Press ◄ / ► to set the lower alarm threshold value and confirm with
MENU/ENTER.
“ max” is displayed.
Press ◄ / ► to set the upper alarm threshold value and confirm with
MENU/ENTER.
Setting the alarm sound
Configuration mode is open, “ ” is displayed.
Press ◄ / ► to enable/disable the alarm sound (“ON” / “OFF”) and
confirm with MENU/ENTER.
Carrying out a reset of the setting menus
Configuration mode is open, “M. RES” (menu reset) is displayed.
Press ◄ / ► to select the required option and confirm with
MENU/ENTER:
o NO: Do not carry out reset.